EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V1

GE LOGIQ V1 Basic Service Manual

GE LOGIQ V1
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #203 background imageLoading...
Page #203 background image
Loading the software
LOGIQ V2/LOGIQ V1 – Basic Service Manual 7-13
5610739-100 English Rev.10
Loading the software
Purpose of this section
This section describes how to reinstall software on
LOGIQ V2/LOGIQ V1.
Customer provided prerequisite
Formatted and labelled media for Images storage.
Formatted and labelled media for Patient Archive and
Presets (User Defined Settings).
Password for the user ADM.
Data Management - moving all images
Move the images to a remote server or to removable media.
For instructions, please see “Disk management” in the User
Manual/User Guide.
CAUTION
An error, or a power loss may occur.
Always backup the Patient Archive and the Presets (System
Configurations) before loading the software!
In order to complete a successful restore of the Patient
Database, as needed after a hard disk replacement, or if all the
content on the hard disk has been erased, the images must be
moved away from LOGIQ V2/LOGIQ V1 before doing backup
of the Patient Database.
Depending on the location set-up, either move the images to a
remote server or to removable media like DVD or CD discs.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals