Disassembly/Re-assembly
LOGIQ V2/LOGIQ V1 – Basic Service Manual 7-7
5610739-100 English Rev.10
Trackball Roller Cleaning
1. Rotate the retainer counterclockwise until it can be removed
from the keyboard.
Figure 7-1. Rotate the retainer
2. Seperate the trackball and the retainer. Wipe off any oil or
dust from the trackball, retainer and the trackball housing
using a cleaner or cotton swab.
3. Assemble the trackball and retainer, then put in into the
housing and rotate it clockwise until its notches are set in
position.
CAUTION
When cleaning, make sure not to spill or spray any liquid into
the trackball housing (keyboard or system).