Completing the setup
LOGIQ V2/LOGIQ V1 – Basic Service Manual 3-17
5610739-100 English Rev.10
Peripheral/Accessories Connector Panel (continued)
NOTE: Accessory equipment connected to the analog and digital
interfaces must be certified according to the respective IEC
standards (e.g. IEC60950 for data processing equipment and
IEC60601-1 for medical equipment). Furthermore, all complete
configurations shall comply with the valid version of the system
standard IEC60601-1. Everybody who connects additional
equipment to the signal input part or signal output part of LOGIQ
V2/LOGIQ V1, configures a medical system, and is therefore
responsible that the Ultrasound system complies with the
requirements of the valid version of IEC60601-1. If in doubt,
consult the technical service department or your local
representative for GE.
Connecting probes
Please refer to User Manual on how to connect/disconnect a
probe.
Powering the system
Please refer to User Manual on how to power the system.