EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V1

GE LOGIQ V1 Basic Service Manual

GE LOGIQ V1
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #150 background imageLoading...
Page #150 background image
General Procedures and Functional Checks
4-28 LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
M Mode Checks
M-Mode Screen Example
Preparations
• Connect one of the probes to the scanner’s left-most probe
connector.
• Turn ON the scanner.
The 2D Mode window is displayed (default mode).
• Press M on the Operator panel to bring up an M-Mode
picture on the screen.
Use the trackball to position the cursor over the required
area of the image.
Adjust the M Mode controls
These Image Controls are used to optimize the M mode picture.
Verify that all the listed controls are working as intended:
• Adjust Horizontal sweep to optimize the display resolution.
Adjust Gain and TGC controls to adjust the range to be
imaged.
Use the Frequency (move to higher frequencies) or the
Frame rate control (move to lower frame rate) to increase
resolution in image.
Use the Frequency (move to lower frequency) to increase
penetration.
Adjust Focus to move the focal point(s) around the region of
interest in the M-Mode display.
Adjust Dynamic range to optimize the useful range of
incoming echoes to the available grey scale.
Adjust Compress and Edge Enhance to further optimize
the display.
Adjust Reject to reduce noise while taking care not to
eliminate significant low-level diagnostic information.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals