Functional checks
LOGIQ V2/LOGIQ V1 – Basic Service Manual 4-37
5610739-100 English Rev.10
Probe/Connectors Check
NOTE: Probes can be connected at any time, whether the unit is ON or
OFF.
Connecting a Probe
1. Place the probe’s carrying case on a stable surface and
open the case.
2. Carefully remove the probe and unwrap the probe cable.
3. Put the porbe in the probe holder.
4. Align the connector with the probe port and carefully push
into place.
5. Lock the probe latch upward.
6. Carefully position the probe cord so it is free to move, but
not resting on the floor.
DO NOT allow the probe head to hang free. Impact to the
probe head could result in irreparable damage.
Table 4-1: Probe and Connectors Checks
Step Task Expected Results
1 Select the appropriate connected probe from the
probe indicators on the Touch Panel.
The probe activates in the currently-selected
operating mode. The probe’s default
settings for the mode and selected exam are
used automatically.
2 Launch the application. To change application
without changing the current probe, press Model on
the Touch Panel.
The selected application starts.
3 Verify there’s no EMI/RFI or artifacts specific to the
probe.
No EMI/RFI or artifacts.
4 Test the probe in each active connector slot. It will display pictorial data each time.
5 Do a leakage test on the probe, See ‘Electrical
safety tests’ on page 10-20 for more information.
It passes the test.
6 Repeat this procedure for all available probes.