General procedures
LOGIQ V2/LOGIQ V1 – Basic Service Manual 4-19
5610739-100 English Rev.10
Archiving Presets to a USB memory device (or DVD-R Disk, SD Card)
(continued)
3. Select the item to Backup.
4. Enter backup destination or browse through the disk to
locate the destination.
5. Select Backup. The backup status for each item is displayed
on the Result column.
Loading Presets from a USB memory device (or DVD-R)
1. Connect the USB memory device or DVD-R with the
archived Presets to the system.
2. Enter Utility -> System -> Backup/Restore. The Backup/
Restore sheet will be shown on the LCD display. See
Figure 4-12 on page 4-18.
3. Select the items needed to be restored.
4. Select Restore. The system performs the restore and
restarts.