General procedures
LOGIQ V2/LOGIQ V1 – Basic Service Manual 4-23
5610739-100 English Rev.10
Clean the Trackball
Dust is often building up behind the ball, so it interferes with the
ball rotation and for optical trackballs the light used for sensing.
To get access for cleaning, you need to remove the ball.
The ball is held in position by the Dust Gasket.
1. Power off the system.
2. Rotate the dust gasket counterclockwise until it can be
removed from the keyboard.
Figure 4-13. Remove the retainer
3. Separate the trackball and the gasket. Wipe off any oil or
dust from the trackball, gasket and the trackball housing
using a cleaner or cotton swab.
4. Assemble the trackball and gasket, then put it into the
housing and rotate it clockwise until its notches are set in
the position.
Test the Trackball
Power up the system and test that the trackball now works as
intended.
CAUTION
When cleaning, make sure not to spill or spray any liquid into
the trackball housing (keyboard or system).

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals