EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V1

GE LOGIQ V1 Basic Service Manual

GE LOGIQ V1
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #303 background imageLoading...
Page #303 background image
Electrical safety tests
LOGIQ V2/LOGIQ V1 – Basic Service Manual 10-27
5610739-100 English Rev.10
Data Sheet for enclosure Source Leakage Current
The test passes when all readings measure less than the value
shown in Table 10-12 on page 10-22. Record all data on the PM
Inspection Certificate.
Table 10-16: Typical Data Sheet for enclosure Source Leakage Current
Unit
Power
Tester
Polarity
Switch
Tester
Neutral
or
Ground
Switch
Test 1
Speaker
Cover
Test 2 Real
Panel Metal
Parts
Optional
Test 3
Optional
Test 4
Enter Name of tested peripheral here:
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals