Care and Maintenance
10-30 LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
Data Sheet for Transducer Source Leakage Current
The test passes when all readings measure less than the values
shown in Table 10-13 on page 10-22. Record all data on the PM
Inspection Certificate.
Table 10-17: Typical Data Sheet For Transducer Source Leakage Current
Transducer Tested:
Unit Power
Tester Power Polarity
Switch
Tester GROUND or
NUETRAL Switch
Measurement
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED