EasyManuals Logo
Home>GE>Medical Equipment>MAC 1200 ST

GE MAC 1200 ST Service Manual

GE MAC 1200 ST
151 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
9-10 MAC 1200 resting ECG analysis system Revision D
2012250-095
Specifications: Basic Information
Mains Operation
Instrument design in protection class I according to IEC 601-1
Rated voltage range: 95 to 240 V
Operating voltage range: 85 to 264 V,
49 to 65 Hz
Rated current: 0.2 ... 0.6 A
Fuse: 2 x T 1.25 A, 5 x 20
Typical power consumption:
Battery charging: 14 W
Maximum power consumption: 29 W
Battery Operation
Type: nickel-cadmium
Rated battery voltage: 18 V
Rated battery capacity: 1.3 Ah
Fully charged battery sufficient for up to 50 12 Lead Mode, 1-page ECGs, if
unit is only switched on to record the ECGs
Battery charge time: approximately 4 hours
(Minimum charge time for one 12 Lead Mode
ECG: 10 minutes)
Battery life: approximately 2 to 3 years, replacement by service only
Lithium battery for built-in clock, battery life: approximately 5 years,
replacement by service only
Operational Readiness
After successful self-test, approximately 20 seconds after power-up
Operating Position
Horizontal

Table of Contents

Other manuals for GE MAC 1200 ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 1200 ST and is the answer not in the manual?

GE MAC 1200 ST Specifications

General IconGeneral
BrandGE
ModelMAC 1200 ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals