1-8 MAC 1200 resting ECG analysis system Revision D
2012250-095
The Basics: Mechanical Components
Mechanical Components
The main mechanical components comprise the upper and lower shell of the MAC
1200 resting ECG analysis system. The lower shell serves as a basic unit to receive
the following assemblies.
Line power input module with fuse
Switching power supply
NC Battery
PCB NC Battery Charge
PCB Control CS_CI
Graphics display
Thermal printing unit
Paper container
The upper shell accommodates the keypad, which is connected to the PCB Control
CS_CI via a 26-pin connector.
The MAC 1200 resting ECG analysis system can be opened by releasing 5 screws
at the bottom of the lower shell, opening the paper flap and lifting up the upper shell
a little bit to remove the keypad cable from the PCB Control CS_CI. The upper shell
then can be removed completely.
To replace the NC Battery, the battery flap at the bottom of the lower shell must first
be unlocked with a screw-driver before it can be removed.
The 15-pin input connector to connect the patient cable and the 9-pin connector for
the RS-232 interface are located directly on the PCB Control CS_CI.

Table of Contents

Other manuals for GE MAC 1200 ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 1200 ST and is the answer not in the manual?

GE MAC 1200 ST Specifications

General IconGeneral
BrandGE
ModelMAC 1200 ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals