Revision D MAC 1200 resting ECG analysis system 2-23
2012250-095
Functional Description: Interfaces to Peripherals
Interfaces to Peripherals
The MAC 1200 resting ECG analysis system has only 3 interfaces for peripherals.
Mains input
Patient input
RS-232 interface
The mains input interface on the device is a 3-pin standard cold appliance socket
connection, which is integrated into the mains input module. Connection to the
mains is effected via a 3-pin power cord with a non-fused earth conductor.
The mains input is a wide range input from 95 Vac ... 240 Vac. The mains input
module contains the two primary fuses.
When replacing these fuses, the following points need to be observed.
Replacement by factory or servicing agent only.
Disconnect mains plug.
Replacement only with the original fuses.
For the patient input, the mechanics of the patient input of the MAC 500 is used, so
the patient cables of the MAC 1000/CardioSmart cannot be used for the MAC
1200 resting ECG analysis system.
A 9-pin sub-D Connector, type Mark II with 4-40 UNC inserts, with a standard
configuration of the signals TXD, RXD and GND is implemented in the
construction of the RS-232 interface.

Table of Contents

Other manuals for GE MAC 1200 ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 1200 ST and is the answer not in the manual?

GE MAC 1200 ST Specifications

General IconGeneral
BrandGE
ModelMAC 1200 ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals