ORTHORALIX 9200 Service Manual
B-8
Gendex Dental Systems 4519-190-71193-7
B.5.3 Installing the z-carriage on to the
column
The Z-carriage is delivered assembled with rotating unit and
tubehead and shipped turned upside down.
The rotating unit and the Z-carriage are connected by a fixing
bracket. The following covers are disassembled in the packing:
tUPQDPWFS
tGSPOUDPWFSXJUINJSSPS
tSFBSDPWFSPGDBTTFUUFVOJUOPUQSFTFOUJOUIF0SUIPSBMJY
DDE versions).
The Z-carriage is heavier than 20 kg so, to handle it
safely, three men are needed, two at the sides and
one at the front. Grasp the handles at the bottom
and at the front of the top Z-carriage frame.
Do not handle the Z-carriage by the patient hand-
grip, the headrest or the tubehead.
NO
NO
To install the Z-carriage on the column, proceed as follows:
1. Rotate the Z-carriage upside down, by turning around the
rear side, and move it close to the base of the column.
(03.0)E
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER

Table of Contents

Other manuals for Gendex ORTHORALIX 9200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gendex ORTHORALIX 9200 and is the answer not in the manual?

Gendex ORTHORALIX 9200 Specifications

General IconGeneral
BrandGendex
ModelORTHORALIX 9200
CategoryMedical Equipment
LanguageEnglish

Related product manuals