ORTHORALIX 9200 Service Manual
B-21
(06.0)EGendex Dental Systems 4519-190-71193-7
B.5.13.2 Orthoralix 9200 DDE/Ceph
Two lead plates are supplied with the system to allow all the fol-
lowing Ceph X-rays formats:
a) one lead plate height 18 cm (H 18 written on the lead plate)
suitable for:
tMBUFSPMBUFSBMIPSJ[POUBMQSPKFDUJPOYDN
b) a second lead plate height 22 cm (H 22 written on the lead
plate) suitable for:
tMBUFSPMBUFSBMWFSUJDBMQSPKFDUJPOYDN
tMBUFSPMBUFSBMNBYQSPKFDUJPOYDN
tBOUFSPQPTUFSJPSWFSUJDBMQSPKFDUJPOYDN
tBOUFSPQPTUFSJPSNBYQSPKFDUJPOYDN
tQPTUFSPBOUFSJPSWFSUJDBMQSPKFDUJPOYDN
tQPTUFSPBOUFSJPSNBYQSPKFDUJPOYDN
Do not install the lead plates on the collimator till all the
ceph alignments are done (see Section C and Section F of
this Service Manual).
These lead plates are already installed on the lead plate guide
supplied with the ceph arm; this lead plate must be installed on
the collimator by following the procedure below:
1. Remove the front cover (a) of the tubehead by unscrewing
the two screws (b) and by disengaging the spring latches
from the rear cover (c).
2. Remove the plate (d) from the collimator by unscrewing the
two screws (e).
3. Install the lead plate guide (f) on the collimator, using the
two screws (g) provided.
If different screws are used instead of the one sup-
plied, it is possible:
tUPEBNBHFUIFUISFBE
tUPIBWFJOUFSGFSFODFCFUXFFOUIFGJYFEBOEUIF
mobile parts of the collimator
Before remounting the tubehead front cover, adjust the X-ray field
of the Ceph formats as described in the Section C.
(03.0)E