ORTHORALIX 9200 Service Manual
B-29
(06.0)EGendex Dental Systems 4519-190-71193-7
B.10 Grounding Connection
Supplemental Installation Instructions
(Orthoralix
®
9200):
Grounding Connection to AC Service in Patient
Care Area
The following method shall be used to ground AC Service to
the Orthoralix 9200 in patient care areas. These instructions
supplement the installation instructions provided in the
Orthoralix 9200 Service Manual, Section B.
Electrical connections to the equipment are made at the base of
the column assembly. Remove hex bolts from left side and front
column covers to obtain access to the electrical service area.
1. Create an opening for AC service to the equipment in the
angled section of the left column side cover at the base of the
column (see Figure 1). Size the opening in accordance with
the type and size of the AC service being provided (e.g. metal
conduit)
Figure 1: Column side cover (w/ indicated hole position), conduit clamp, strain relief, terminal strip
2. Route AC service through opening in the column side cover.
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER

Table of Contents

Other manuals for Gendex ORTHORALIX 9200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gendex ORTHORALIX 9200 and is the answer not in the manual?

Gendex ORTHORALIX 9200 Specifications

General IconGeneral
BrandGendex
ModelORTHORALIX 9200
CategoryMedical Equipment
LanguageEnglish

Related product manuals