5-6
CONFINED SPACE MODEUSER HANDBOOK
Note: When a manual log is taken, the term ‘LOG’ is inversed on
the display for one (1) second to provide a visual conrmation
that the log has been captured.
5.5.6 Menu / Off
To re-select the Mode Menu:
• Press and hold both the LH button and RH button .
When the menu appears on the screen, release the buttons,
otherwise the instrument will proceed into the switch OFF
process and will switch OFF after a further three (3) seconds.
Note: In Conned Space Mode, to prevent inadvertently switching
instrument OFF or changing mode while alarms are active,
the user must press and hold both the LH and RH
buttons simultaneously for an additional ve (5) seconds
before the mode menu appears.
5.6 CSM PUMP
The pump will run continuously for safety reasons.
Flow Fault will be agged and fault LED illuminated.
To clear the ow fault:
• Press and hold the RH button .
5.7 CSM ALARMS
Refer to Chapter 11.
5.8 CSM LOGGING
Automatic datalogging is active for all ranges available. Timed logs
will be recorded every minute (default) or as per conguration.
Manual Logging is also available. This will log all gas concentrations
at that instant.