5-6
CONFINED SPACE MODEUSER HANDBOOK
Note: When a manual log is taken, the term ‘LOG’ is inversed on
the display for one (1) second to provide a visual conrmation
that the log has been captured.
5.5.6 Menu / Off
To re-select the Mode Menu:
Press and hold both the LH button and RH button .
When the menu appears on the screen, release the buttons,
otherwise the instrument will proceed into the switch OFF
process and will switch OFF after a further three (3) seconds.
Note: In Conned Space Mode, to prevent inadvertently switching
instrument OFF or changing mode while alarms are active,
the user must press and hold both the LH and RH
buttons simultaneously for an additional ve (5) seconds
before the mode menu appears.
5.6 CSM PUMP
The pump will run continuously for safety reasons.
Flow Fault will be agged and fault LED illuminated.
To clear the ow fault:
Press and hold the RH button .
5.7 CSM ALARMS
Refer to Chapter 11.
5.8 CSM LOGGING
Automatic datalogging is active for all ranges available. Timed logs
will be recorded every minute (default) or as per conguration.
Manual Logging is also available. This will log all gas concentrations
at that instant.

Table of Contents

Other manuals for GMI GT series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GMI GT series and is the answer not in the manual?

GMI GT series Specifications

General IconGeneral
BrandGMI
ModelGT series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals