CP : control panel, FC : frequency converter, SB : supply board, CB : cooling board, CC : control cable, LL : lid locking, BC : braking chopper,
BR : brake resistor, MR : mains reset, EC : error cause, ES : error consequence, ER : error remedy, M : measurements, ECR : error-code reset
15/51
4.2. Description and elimination of errors
TACHO - ERROR 01
EC During centrifugation the speedometer pulses are interrupted.
ES The rotor slows down until it stops.
After the rotor stops, there is a DC braking for 30 sec.
An MR during slowing-down causes a DC braking for 3 min.
After the DC braking, the ”open the lid” release takes place.
Further cooling to NOMINAL temperature.
ER
Speed sensor (speedometer) defective or loose contact on plug.
Measure speedometer pulses on plug X4 / SB, (pin 4 - pin 2).
CC to CP, or CC to FC is defective.
SB or CP or FC is defective.
M Also see at SB-X4, CP-X1 (PIN 14 and FC-S501 (PIN 8).
ECR Open the lid. Turn the rotor by hand and perform an MR while the
rotor is turning.
TACHO - ERROR 02
EC There are no speedometer pulses on the CP after startup.
ES The rotor slows down until it stops.
After the rotor stops, there is a DC braking for 30 sec.
An MR during slowing-down causes a DC braking for 3 min.
After the DC braking, the ”open the lid” release takes place.
Further cooling to NOMINAL temperature.
ER
Startup took place without the rotor.
Motor not connected.
Motor is defective.
Speed sensor (speedometer) defective, or loose contact on plug.
Measure speedometer pulses on plug X4 / SB, (pin 4 - pin 2).
CC to CP, or CC to FC is defective.
No release signal to FC.
SB or CP or FC is defective.
M Also see at SB-X4, CP-X1 (PIN 14 and FC-S501 PIN 8).
ECR Open the lid. Turn the rotor by hand and perform an MR while the
rotor is turning.