NOTICE
Damage to the device electronics due to incorrect voltage or
frequency at the device circuit breaker.
− Operate the device with the correct mains voltage and
mains frequency.
The value can be found in the technical data and on the
rating plate.
NOTICE
Damage to the device and samples due to premature pro-
gram termination.
Premature program termination is caused by power failure,
switching off during the program or pulling out the mains
plug.
− Do not switch off the device while the program is run-
ning.
−
Do not trigger the emergency release on the device
while the program is running.
− Do not pull out the mains plug while the program is
running.
3
Device overview
3.1
Technical data
Manufacturer Andreas Hettich GmbH & Co.KG,
D-78532 Tuttlingen
Model MIKRO 220
Type 2200 2200-01
Mains voltage (
±10%)
200-240V1
~
110-127V1
~
Mains frequency 50-60 Hz 50-60 Hz
power consumption 510 VA 510 VA
Power consumption 2.5A 5.3A
max. capacity 60x2.0ml
max. permissible density 1.2kg/dm
³
max. speed (RPM) 18000
max. acceleration (RCF) 31514
max. kinetic energy 8700 Nm
Obligation to perform
checks
(DGUV Rules 100-500)
(valid only in Germany)
No