Troubleshooting
Greasing the motor shaft
Accessories with a limited
service life
Replacing centrifuge tubes
1. Remove accessories.
2. Clean the motor shaft.
3. Remove any detergent residues with a damp cloth after using deter-
gents.
4. Grease the motor shaft with Hettich Tubenfett 4051.
5. Excess grease in the centrifuging chamber must be removed.
The use of certain accessories is time-limited. For safety reasons, the
accessories must no longer be used when either the maximum number
of permissible run cycles marked on them or the expiry date marked on
them has been reached.
The maximum permissible number of run cycles or the expiry date can
be seen marked on the accessories.
The centrifuge is equipped with a cycle counter.
CAUTION
Risk of injury from broken glass.
Broken glass may cause glass splinters and contaminated
liquids to be found inside the centrifuge.
Wear cut-resistant gloves.
Wear protective goggles and a face mask.
Broken parts of the tube, glass splinters and spilled centrifuge material
must be removed completely in the event of leakage or if a centrifuge
tube breaks. Glass splinters that are not removed will cause further glass
breakage.
The rubber inserts and the plastic sleeves of the rotors must be replaced
after a glass breakage.
Disinfection must be carried out if the material is infectious.
9
Troubleshooting
9.1
Fault description
Customer service must be notied if the fault cannot be rectied based on
the fault table. State the centrifuge type and serial number. Both numbers
can be seen on the type plate of the centrifuge.
* Error number does not appear on the display.
Fault description Cause Remedy
no display No power. Overcurrent protec-
tion fuse has tripped.
Check the supply voltage.
Set the mains switch to
[I]
.
TACHO - ERROR 1, 2, 96 Tacho defective. Motor, elec-
tronics defective.
Open the lid.
Set the mains switch to
[0]
.
Wait at least 10 seconds.
Turn the rotor vigorously by hand.
Set the mains switch to
[I]
. The rotor
must rotate while switching on.

Table of Contents

Other manuals for Hettich MIKRO 220