AB2200en_SARev.: 04 / 11.2023
Set
Parameter query
2. Press the
[STOP/OPEN]
button.
🡆 Precooling of the rotor is terminated.
Ramp-down takes place with the selected brake level.
The brake level is displayed.
The precooling speed is adjustable from 500RPM up to the maximum rotor
speed in increments of 10 RPM. It is preset to 10000RPM.
â– The rotor is stationary.
â–
The lid is open.
1. Press and hold the
[Cooling]
button.
🡆
‘t/min = X ’
is displayed after 8 seconds.
2. Press the
[Cooling]
button.
🡆 Precooling speed
‘RPM = XXXX’
is displayed.
3. Use the
[Rotary knob]
to set the precooling speed.
4. Press the
[START/IMPULS]
button.
🡆 The setting is stored.
‘*** ok ***’
is displayed brieîž…y.
5.
Press the
[STOP/OPEN]
button twice or wait 8 seconds to exit the
menu.
7.5
Machine Menu
7.5.1 Querying system information
The rotor is stationary.
1.
Press and hold the
[SELECT]
button for 8 seconds.
🡆
‘SOUND/BELL’
is displayed.
2.
Press the
[SELECT]
button repeatedly until
‘FU/CCI – S.’
is displayed.
Program version for the frequency inverter
3.
Press the
[SELECT]
button repeatedly until
‘HOURS’
is displayed.
Internal operating hours (the time during which the centrifuge was
switched on)
4.
Turn to the right with the
[Rotary knob]
.
🡆
‘STARTS’
is displayed.
Number of centrifugation runs
5.
Turn to the right with the
[Rotary knob]
.
🡆
‘ROTORCHG1’
is displayed.
Internal operating hour of the last rotor change
6.
Turn to the right with the
[Rotary knob]
.
🡆
‘ROTORCHG2’
is displayed.
Internal operating hour of the penultimate rotor change
7.
Turn to the right with the
[Rotary knob]
.
🡆
‘OPhoursCHG’
is displayed.
Internal operating hour of the last operating hours change
8.
Turn to the right with the
[Rotary knob]
.
🡆
‘IMBALCHG’
is displayed.
Internal operating hour of the last imbalance cut-off change