VersaCare® Bed User Manual (161956 REV 2) 5
To Activate
1. Make sure the bed is plugged into AC power or the battery is enabled.
NOTE:
To enable the battery, press any of the caregiver controls except the Lockout control.
2. Pull the CPR control handle with one hand.
3. Hold the handle until the head and knee sections come to a stop in the flat position, if
not currently in the flat position.
4. Continue to hold the CPR control handle until the desired angle is reached.
The head section actuator is automatically re-enabled after the CPR control handle is
released.
Caregiver Siderail Controls
The Caregiver Siderail controls are located on the outside of each head-end siderail.
There are two sets of Caregiver Siderail controls. The first set is mounted on the outside of
both siderails and control the bed position functions. The second set, for the optional bed
functions, is mounted on a flip-up control pod in the head-end siderails. The second set of
controls is for the scale, treatment/therapy surface, and the Bed Exit Alarm System.
Instruct visitors not to attempt operation of caregiver controls. They may assist the patient
with patient controls.
Enable
The Enable control is located on the optional flip-up control pod. The Enable
control deters unauthorized operation of certain caregiver controls. With the
exception of the Weigh control, the Enable control must be activated before
the caregiver controls on the pod will operate. When activated, the Enable
indicator stays on for 60 seconds. During this time, the caregiver can use any caregiver
controls on the pod.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom VersaCare Bed and is the answer not in the manual?

Hill-Rom VersaCare Bed Specifications

General IconGeneral
TypeHospital Bed
ApplicationAcute Care
Point-of-Care ControlsYes
X-Ray Cassette SleeveOptional
CPR ReleaseYes
Battery BackupYes
Bed Width36 inches
Safe Working Load500 lb (227 kg)
Side Rail OptionsFull