VersaCare® Bed User Manual (161956 REV 2) 27
30° Head Angle Alarm
The Head Angle Alarm control is on the caregiver pod next to the display.
When set, if the head section goes below 30°, these will occur:
• The display will flash five times.
• An audible alarm will come on.
• The alarm indicator will flash.
The display on the caregiver pod continuously shows the angle
of the head section. Whenever the head section goes below 30°,
the display flashes five times.
Set the Alarm
1. Raise the head section to the applicable position above 30°.
2. Press the Enable control.
3. Press the Alarm control. The alarm indicator will come on.
When the bed operates on battery power the display will be off. However, if the alarm is
set and the head section goes below 30°, an audible alarm will come on and the alarm
indicator will flash.
Turn Off the Alarm—raise the head section above 30°.
Deactivate the Alarm
1. Press the Enable control.
2. Press the Alarm control. The alarm indicator will go off.
Line Manager
Do not use the Line Manager for ventilator circuits. To do
so could cause patient injury.
Do not hang cords on the Line Manager. To do so could
cause injury or equipment damage.
When you use the Line Manager, make sure the lines are not pinched
or kinked and there is sufficient slack in the lines for bed articulations and patient
movement. Failure to do so could cause injury or equipment damage.
Failure to keep aseptic lines separate from non-aseptic lines may cause cross
contamination.