3. Wipe down the surface with chlorine bleach (50 ppm to 150 ppm) or mild detergent
and warm water followed by an approved intermediate level disinfectant, such as
CSI disinfectant cleaner.
NOTE:
2.5 oz of bleach per 10 gal of water is approximately 100 ppm of available chlorine.
4. Let the bleach or disinfectant remain in contact with the surface as instructed in the
manufacturers instructions.
5. Remove the bleach or disinfectant, and rinse with warm water.
6. Let the mattress and/or coverlet completely air dry.
7. If the sleep surface was removed, install it on the bed. Go to “To Install The Sleep
Surfaces” on page 21.
Machine Wash the AMT Coverlet
1. Remove the coverlet.
2. Machine-wash the coverlet with chlorine bleach (50 ppm to 150 ppm) or detergent
and an effective intermediate level disinfectant, such as CSI disinfectant cleaner.
(For customers in the US, the disinfectant should be registered with the
Environmental Protection Agency.)
NOTE:
2.5 oz of bleach per 10 gal of water is approximately 100 ppm of available chlorine.
Use the bleach or disinfectant as instructed in the manufacturers instructions.
To determine the amount of bleach or disinfectant to use, determine the amount
of water in the washer, and follow the manufacturers dilution instructions.
During the wash cycle, soak the coverlet in the disinfectant or bleach.
Let the coverlet rinse thoroughly in clean water.
CAUTION:
Do not use high temperatures to dry the coverlet. Air dry or select a low or non-
heat dry cycle such as air fluff. High temperatures could destroy the coating that
makes the coverlet waterproof yet breathable.
3. Use the lowest temperature setting of the dryer to dry the coverlet. Do not use high
temperatures.
4. Put the coverlet on the mattress.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom VersaCare Bed and is the answer not in the manual?

Hill-Rom VersaCare Bed Specifications

General IconGeneral
TypeHospital Bed
ApplicationAcute Care
Point-of-Care ControlsYes
X-Ray Cassette SleeveOptional
CPR ReleaseYes
Battery BackupYes
Bed Width36 inches
Safe Working Load500 lb (227 kg)
Side Rail OptionsFull

Related product manuals