Point Gas Monitor Technical Manual
29
Monitor Operation
Network Model Alarm Menu
CAUTION: Only qualified, knowledgeable personnel should use the
programming functions of this monitor. Factory settings conform to specific
standards (see Specifications). Any changes made to Alarm Levels may
affect manufacturer’s stated standards compliance.
The Alarm menu allows users to define different alarm levels than the
factory settings. Press enter to access the Alarm menu options. If there
is a remote sensor connected, scroll to select Z1 (built-in sensor) or Z2
(remote sensor, not available on Network Model).
* Menu *
Alarm
Alarm
Z1
• Use the arrow keys to scroll to the Alarm A, B, or C option and
press enter to select. The screen displays the default alarm level.
Alarm
A
• Use the arrow keys to increase or decrease the value to the
desired level and press enter.
Alarm A
0025 PPM
When alarm settings are modified by the customer, the new values are
stored in the main device and do not need to be modified again when a
sensor is replaced.
WARNING: If a different type of gas sensor catridge is installed, the
previously stored values will no longer be valid and must be corrected.
• The next screen displays the hysteresis value. Use the arrow keys
to increment or decrement the value and press enter.
Hyst A
-05