Getting Started
Copyright ©ITECH Electronic Co., Ltd. 37
NOTE
After entering the menu interface, the knob can also be used to scroll pages to view menu
items.
4.4 Input On/Off Control
The [On/Off] key is used to turn the input on or off under normal
circumstances. Even if the instrument is in control by PC or the keyboard is
locked, the [On/Off] is still valid.
The [On/Off] key light is off and turning the input off does not place the
instrument in a safe state. Hazardous voltages may be present on all input
and guard terminals. Putting the equipment into an input-off state does not
guarantee that the inputs are powered off if a hardware or software fault
occurs. See the cautions about connecting the test lines before connecting
test lines.
You can press the [On/Off] key on the front panel to control the input status of
the instrument. If the [On/Off] key light is on, indicates that the input is turned
on. The VFD displays the meter value such as voltage, current, power and so
on. If the [On/Off] key light is off, indicates that the input is turned off. The VFD
displays that the load state is OFF.

Table of Contents

Other manuals for ITech IT8200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8200 Series and is the answer not in the manual?

ITech IT8200 Series Specifications

General IconGeneral
BrandITech
ModelIT8200 Series
CategoryMeasuring Instruments
LanguageEnglish