System-Related Functions
Copyright ©ITECH Electronic Co., Ltd. 75
2. Press the composite keys [Shift]+ (System) on the front panel to
enter the system menu.
3. Select General menu.
4. Set the Parallel Mode, set them to single.
5. Power off the three instruments and turn off the main switch of the AC
distribution box.
6. Remove the cables connection of the System Bus and output terminals
between three units.
7. Power on the three instruments separately.
After the instrument is restarted, the screen shows that the instrument is working
in single mode.
6.12 Remote Measurement Function
The IT8200 series instrument supports two connection methods: Local
measurement and Remote sensing. The remote sensing is used for maximizing
measurement accuracy. (Refer to 2.4 Connecting Test Lines).
The procedures to set the menu item are as follows.
1. Press the composite keys [Shift] + (System) on the front panel to
enter the system menu.
2. Press the up/down key or rotate the knob to select Sense compensation
and press [Enter].
⚫ Local: Default value, indicates turn the sense function off.
⚫ Remote: Indicates turn the sense function on.
3. After the parameter settings are complete, press [Enter].
6.13 Digital I/O Function
This series load supports digital I/O function. The user can realize logic control
over high and low level input or output by related configurations in the system
menu.
Pins Introduction
Different I/O implements different functions. The detailed functions description
are shown in the figure below:
Digital-IO