Operation and Application
Copyright ©ITECH Electronic Co., Ltd. 43
Under the config menu interface, the user can select Const Mode item and set
the mode to CP.
Const Mode: load mode
Unit PF: Unity power factor
On: enable the UPF mode, at this situation, the current waveform changes
with the voltage waveform and this mode causes power factor to be as close
as possible to unity, useful when the input voltage is non-sinusoidal.
Off: disable the UPF mode, at this situation, the current waveform is not
affected by voltage Waveform, and the current waveform is subject to
defined in the config menu, the actual power factor is affected by voltage
Waveform, current Waveform, and phase shift.
Power setting: Constant power value
Current AC Slew Rate: current slope
Waveform (phase shift range: -90.0~90.0): select the waveform. Under CC
mode, and Unit PF is off, Waveform can be select.
Crest Factor: CF range is 1.414~5, and the range is limited by peak value.
Phase shift: Phase shift of voltage and current. Phase shift range within -
90°~90°. If programmed is positive it will be a leading power factor. If
programmed is negative it will be a lagging power factor.
Programming the load value (CP)
The user can press [Set] and set the input resistance value in the main interface
of CP mode. Directly set the present value through the knob or press numeric
keys to input the value. In case of wrong input by numeric keys, press
to delete the present input.
5.3.4 Constant Apparent Power Mode (CS)
In the constant apparent power mode, the constant power is required for the
electronic load, and the current sinks according to the set apparent power, as
shown below. If the input voltage increases, the input current will decrease to
keep the power S (=V * I) at the set value.

Table of Contents

Other manuals for ITech IT8200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8200 Series and is the answer not in the manual?

ITech IT8200 Series Specifications

General IconGeneral
BrandITech
ModelIT8200 Series
CategoryMeasuring Instruments
LanguageEnglish