CALibration Subsystem
Copyright © Itech Electronic Co., Ltd. 105
Chapter13 CALibration Subsystem
Calibration commands let you:
Enable and disable the calibration mode
Calibrate the input functions, current monitor offset and gain, and store new
calibration constants in nonvolatile memory.
CALibrate:SECure[:STATe]
This command enables and disables calibration mode. The calibration mode
must be enabled before the load will accept any other calibration commands.
The first parameter specifies the enabled or disabled state. The second
parameter is the password. It is required if the calibration mode is being
enabled and the existing password is not 0. If the password is not entered or is
incorrect, an error is generated and the calibration mode remains disabled. The
query statement returns only the state, not the password. Whenever the
calibration state is changed from enabled to disabled, any new calibration
constants are lost unless they have been stored with CALibrate:SAVE.
Command Syntax
CALibrate:SECure[:STATe] <bool> [,<SRD>]
Parameters
0 | 1 | OFF | ON [,<password>]
Returned Parameters
ON
Examples
CAL:SEC 0, N3301A CAL:SEC ON
Query Syntax
CALibrate:SECure[:STATe]?
Returned Parameters
<NR1>
Related Commands
CAL:SAVE CAL:INIT