IEEE-488 Command Reference
Copyright © Itech Electronic Co., Ltd. 29
NOTE: The device state stored in location 0 is automatically recalled at power
turn-on.
Command Syntax
*RCL <NRf>
Parameters
0 to 9
Example
*RCL 3
Related Commands
*PSC
*RST
*SAV
*RST
This command resets ALL channels of the electronic load to the following
factory-defined states:
INP OFF INP:SHOR OFF
TRAN OFF FUNC:MODE FIX
FUNC CURR CURR:RANG MAX
CURR MIN CURR:SLEW MAX
CURR:TRAN:ALEV MAX CURR:TRAN:AWID MIN
CURR:TRAN:BLEV MIN CURR:TRAN:BWID MIN
CURR:TRAN:MODE CONT VOLT:RANG MAX
VOLT MAX VOLT:SLEW MAX
VOLT:TRAN:ALEV MAX VOLT:TRAN:AWID MIN
VOLT:TRAN:BLEV MIN VOLT:TRAN:BWID MIN
VOLT:TRAN:MODE CONT RES:RANG MAX
RES MAX RES:SLEW MAX
RES:TRAN:ALEV MAX RES:TRAN:AWID MIN
RES:TRAN:BLEV MIN RES:TRAN:BWID MIN
RES:TRAN:MODE CONT CURR:PROT:STAT OFF