Function and Features
Copyright © ITECH Electronic Co., Ltd. 24
Constant Resistance
High=130.0V
4. Set minimum voltage value and press [Enter] key for confirmation.
Constant Resistance
Low=0.000V
5. Complete parameter setting.
10.0000V 0.0000A
0.00W CR=2.000Ω
NOTE
If the above method is for editing auto test step (as mentioned below), constant
resistance range can also be set.
3.2.4 Constant power operation mode (CW)
Under CW mode, the electronic load will consume a constant power, as shown
below. If input voltage rises, the input current decreases and power P (= V * I)
will maintain at setting power. As shown in Fig. 3-4:
Fig. 3-4 Voltage-Current Relation Schema under CW Mode
Under CW mode, the electronic load provides three ways to modify constant
power.
Under CW mode, rotate the pulse knob to change constant power value.
Under CW mode, directly input numeric key and press [Enter] key for
confirmation to change constant power value.
Under CW mode, move the cursor with Left/Right Key and press Left/Right
Key to adjust value at corresponding position.
Operation steps
1. Press [CW] key. Press [Shift] + [CV] to enter parameter setting screen.
Constant Power
Range=400.00W
2. Set maximum working power value and press [Enter] key.
Constant Power
Input
voltage
Load current
V
I
Set power
V2
V3
I2
I3
CP mode

Table of Contents

Other manuals for ITech IT8800 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8800 Series and is the answer not in the manual?

ITech IT8800 Series Specifications

General IconGeneral
BrandITech
ModelIT8800 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals