2260B Series User Manual
108
4. Press the Output key. The voltage
can now be controlled with the
External resistance.
Note
Ensure the resistor(s) and cables used exceed the
isolation voltage of the power supply. For example:
insulation tubes with a withstand voltage higher
than the power supply can be used.
When choosing an external resistor ensure the
resistor can withstand a high degree of heat.
Note
CV and CC Slew Rate Priority are disabled for V-I
mode (F-03) when using external resistance
control. See the normal function settings on page
87.
External Resistance Control of Current Output
Background
External resistance control of the current output is
accomplished using the MIL-26 connector on the
rear panel. A resistance of 0k~10k is used to
control the full scale current of the instrument.
The output current (0 to full scale) can be
controlled with the external resistance going up
(Ext-R ) 0k~10k(10k = Vo,max) or down
(Ext-R ) 10k~0k(10k = 0).
For 0k~10k: Output current = full scale current
× (external resistance/10)
For 10k~0k: Output current = full scale current
× ([10-external resistance]/10)

Table of Contents

Other manuals for Keithley 2260B-80-13

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2260B-80-13 and is the answer not in the manual?

Keithley 2260B-80-13 Specifications

General IconGeneral
BrandKeithley
Model2260B-80-13
CategoryPower Supply
LanguageEnglish

Related product manuals