OPERATION
33
Wire Gauge Considerations
Background
Before connecting the output terminals to a
load, the wire gauge of the cables should be
considered.
It is essential that the current capacity of the
load cables is adequate. The rating of the cables
must equal or exceed the maximum current
rated output of the instrument.
Recommended
wire gauge
Wire Gauge
Maximum Current
20
2.5A
18
4A
16
6A
14
10A
12
16A
10
21A
8
36A
6
61A
4
97A