EasyManuals Logo
Home>Keithley>Power Supply>2260B-80-13

Keithley 2260B-80-13 User Manual

Keithley 2260B-80-13
153 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #51 background imageLoading...
Page #51 background image
2260B Series User Manual
48
Set to CC Mode
When setting the power supply to constant current mode, a voltage
limit must also be set to determine the crossover point. When the
voltage exceeds the crossover point, the mode switches to CV mode.
For details about CC operation, see page 18. CC and CV mode have
two selectable slew rates: High Speed Priority and Slew Rate
Priority. High Speed Priority will use the fastest slew rate for the
instrument while Slew Rate Priority will use a user-configured slew
rate.
Background
Before setting the power supply to
CC mode, ensure:
The output is off.
The load is connected.
Steps
1. Press the Function key. The
Function key will light up.
2. The display should show F-
01 on the top and the
configuration setting for
F-01 on the bottom.
3. Rotate the Voltage knob to change
the F setting to F-03 (V-I Mode
Slew Rate Select).
Voltage
4. Use the Current knob to set the F-
03 setting.
Set F-03 to 1 (CC High Speed
Priority) or 3 (CC Slew Rate
Priority) and save.
Current

Table of Contents

Other manuals for Keithley 2260B-80-13

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2260B-80-13 and is the answer not in the manual?

Keithley 2260B-80-13 Specifications

General IconGeneral
BrandKeithley
Model2260B-80-13
CategoryPower Supply
LanguageEnglish

Related product manuals