CONFIGURATION
89
Rising Current
Slew Rate
Sets the rising current slew rate. Only
applicable if V-I Mode is set to CC Slew Rate
Priority.
F-06
0.01A/s~72.00A/s (2260B-30-36)
0.1A/s~144.0A/s (2260B-30-72)
0.01A/s~27.00A/s (2260B-80-13)
0.01A/s~54.00A/s (2260B-80-27)
Falling Current
Slew Rate
Sets the falling current slew rate. Only
applicable if V-I Mode is set to CC Slew Rate
Priority.
F-07
0.01A/s~72.00A/s (2260B-30-36)
0.1A/s~144.0A/s (2260B-30-72)
0.01A/s~27.00A/s (2260B-80-13)
0.01A/s~54.00A/s (2260B-80-27)
Internal
Resistance
Settings
Sets the internal resistance of the power supply.
F-08
0.000~0.833 (2260B-30-36)
0.000 ~0.417 (2260B-30-72)
0.000 ~5.926 (2260B-80-13)
0.000 ~2.963 (2260B-80-27)
Bleeder Control
Bleeder control turns ON/OFF the bleeder
resistor. Bleeder resistors discharge the filter
capacitors after power is turned off as a safety
measure.
F-09
0 = OFF, 1 = ON
Buzzer ON/OFF
Turns the buzzer sound on or off. The buzzer is
associated with alarm sounds and keypad entry
sounds.
F-10
0 = ON, 1 = OFF

Table of Contents

Other manuals for Keithley 2260B-80-13

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2260B-80-13 and is the answer not in the manual?

Keithley 2260B-80-13 Specifications

General IconGeneral
BrandKeithley
Model2260B-80-13
CategoryPower Supply
LanguageEnglish

Related product manuals