11-8 DISPlay, FORMat, and SYSTem
Command notes (SCPI commands — system)
SYSTem:POSetup <name> Program power-on defaults
Parameters <name> = RST Power-up to *RST defaults
SAV0 Power-up to setup stored in memory location 0
SAV1 Power-up to setup stored in memory location 1
SAV2 Power-up to setup stored in memory location 2
SAV3 Power-up to setup stored in memory location 3
SAV4 Power-up to setup stored in memory location 4
With RST selected, the power supply powers up to the *RST default conditions. Default
conditions
are listed in the SCPI tables.
NOTE The Model 2306-PJ
, 2306-VS, and 2302-PJ have only SAV0-SAV2 memory locations.
With SAV0-4 specified, the power supply powers-on to the setup that is saved in the specified
memory location using the *SA
V command (Section 9). Note that the instrument will power up
with the output OFF.
SYSTem:AZERo:STATe <b> Enable or disable auto zero.
This command enables (ON) or disables (OFF) auto
zero, which is normally part of the
measurement process used to null internal offsets. Turning off auto zero increases measurement
speed but at the expense of accuracy. To minimize drift when operating with auto zero off for
extended periods, briefly turn on auto zero periodically.
SYSTem:MEP[:STATe] <b> Select GPIB protocol.
This command selects the 488.1 (OFF) or SCPI (ON) GPIB protocol. The 488.1 protocol can
be
used to speed up operation over the GPIB. See Appendix for details.
SYSTem:TRIGger:TALK:BOTH <b> Enable/disable 2-reading trigger on talk.
This command enables and disables 2-reading trigger on talk. See A
ppendix for details.
SYSTem:TRIGger:CONTinuous <b> Enable/disable trigger continuous mode.
This command enables and disables the trigger continuous mode. See Appen
dix for details.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com