D-2 Calibration Program
Introduction
This appendix includes a calibration program written in BASIC to help you in calibrating the
Model 2306. Refer to Section 14 for more details on calibration procedures, equipment, and
connections. Appendix C covers calibration commands in detail.
Computer hardware requirements
The following computer hardware is required to run the calibration programs:
• IBM PC compatible computer.
• Keithley KPC-488.2 or KPC-488.2AT, or
CEC PC-488 IEEE-488 interface for the
computer.
• Two shielded IEEE-488 bus cables (Keithley Model 7007).
Software requirements
To use the calibration program, you will need the following computer software:
• Microsoft QBasic (supplied with MS-DOS 5.0
or later).
• MS-DOS version 5.0 or later.
• HP-style Universal Language Driver, CECHP.EXE (supplied with Keithley and CEC
interf
ace cards listed above).
Calibration equipment
The following calibration equipment is required:
• Keithley Model 2001 Digital Multimeter
•4Ω, 0.1%, 100W re
sistor
•3kΩ, 0.1%,
0.25W resistor
Refer to Se
ction 14 for detailed equipment specifications as well as details on test connec-
tions.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176