13-6 Performance Verification
• Do not connect test equipment to the Model 2306 through a scanner, multiplexer, or
other switching equipment.
• Be sure that the Model 2306 output is turned on before making measurements.
• Allow the unit output signal to settle before making a measurement.
Output voltage accuracy
Follow the steps below to verify that Model 2306 output voltage accuracy is within specified
limits. This test involves setting the output voltage to specific values and measuring the voltages
with a precision digital multimeter.
1. With the power supply’s output off, connect the digital multimeter to the Model 2306
OUTPUT
#1 SOURCE and SENSE terminals, as shown in Figure 13-1. Be sure to
observe proper polarity (SOURCE + and SENSE + to INPUT HI; SOURCE - and
SENSE - to INPUT LO).
Figure 13-1
Connections for voltage verification tests
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
LINE RATING
100-120VAC, 200-240VAC
50, 60 HZ 165VA MAX
LINE FUSE
SLOWBLOW
2.0A, 250V
IEEE-488
(ENTER IEEE ADDRESS
FROM FRONT PANEL MENU)
REMOTE
DISPLAY
OPTION
+++
____
+
SOURCE SENSE
DVM IN
SOURCE
OUTPUT #1
ISOLATION FROM EARTH: 22 VOLTS MAX.
RELAY
CONTROL
24VDC MAX
CAT I
DVM IN
+30 VDC MAX
NEXT
PREV
POWER
DISPLAY
2001 MULTIMETER
DCV ACV DCI ACI Ω2 Ω4
FREQ TEMP
REL TRIG
INFO LOCAL EXIT ENTER
RANGE
RANGE
!
FR
500V
PEAK
FRONT/REAR
2A 250V
AMPS
CAL
STORE RECALL
CHAN SCAN
FILTER MATH
CONFIG MENU
HI
INPUT
LO
SENSE
Ω 4 WIRE
INPUTS
350V
PEAK
1100V
PEAK
AUTO
Model 2001 DMM
Model 2306
Input LO
Input HI
Source -
OUTPUT #1 connections
shown. Use OUTPUT #2
for Charger Channel.
Sense +
Sense -
Source +
+++
____
+
SOURCE SENSE
DVM IN
SOURCE
OUTPUT #2
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2302

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2302 and is the answer not in the manual?

Keithley 2302 Specifications

General IconGeneral
BrandKeithley
Model2302
CategoryPower Supply
LanguageEnglish

Related product manuals