7: Scanning resistors using 4W measurement DAQ6510 Data Acquisition / Multimeter System
7-8 DAQ6510-900-01Rev. A / April 2018
Test results
The table below compares the values of scanned DUTs using 4-wire and 2-wire measurement
methods with the six resistors listed in the Instrument and Device Connection section.
The data provided in each cell of the table is the average of 100 scans.
Nominal
value (Ω)
Four-wire measurement
with offset compensation
(Ω)
Four-wire measurement
with NO offset
compensation (Ω)
Two-wire measurement
(Ω)
2.2 2.20413 2.20588 2.51011