Section
6: Scanning low-level DCV DAQ6510 Data Acquisition / Multimeter System
User's Manual
6-4 DAQ6510-900-01Rev. A / April 2018
Using the front panel
To set up the application from the front panel:
1. Press the POWER switch on the front panel to turn on the instrument.
2. Select the REAR terminals.
3. Press the MENU key.
4. Under Channel, select Scan.
5. Select the + button to add a group of channels, select channels 101 to 106, and select OK.
6. Select DC Voltage.
7. On the Settings tab, set NPLC to 5.
8. Set Auto Zero to On.
9. On the Scan tab, set Scan Count to 10.
10. Select Start at the bottom of the left pane.
To watch active readings during scan:
1. Press the HOME key.
2. Select the arrow to the right of Watch Channel.
3. Select the channels of interest and select OK to accept.
To save the scanned measurements to a USB flash drive:
1. Select the MENU key.
2. Under the Measure column, choose Reading Buffers.
3. Insert a USB flash drive into the DAQ6510. Select Save to USB at the bottom of the pane.
4. Select OK.
Using SCPI commands
This sequence of SCPI commands executes a DCV scan on channels 101 to 106.
You may need to make changes so that this code will run in your programming environment. In the
table, the SCPI commands have a light gray background. The light green shaded code represents
pseudocode that will vary depending on the programming environments you use.

Table of Contents

Other manuals for Keithley DAQ6510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DAQ6510 and is the answer not in the manual?

Keithley DAQ6510 Specifications

General IconGeneral
BrandKeithley
ModelDAQ6510
CategoryMultimeter
LanguageEnglish

Related product manuals