6: Measuring 4-wire resistance with offset compensation Model DMM7510 7½ Digit Multimeter
6-4 DMM7510-900-01 Rev. B / May 2015
Using the front panel
To set up the application from the front panel:
1. Press the POWER button on the front panel to turn on the instrument.
2. On the FUNCTIONS swipe screen, select 4W
to select the 4-wire resistance measure function.
3. Press the MENU key.
4. Under Measure, select Settings.
5. Set Offset Compensation to On.
6. Press the HOME key.
The measurement readings are displayed in the top area of the Home screen.
Using SCPI commands
This sequence of SCPI commands measures the resistance of a device or component. The device
under test is connected to the Model DMM7510 and controlled remotely.
You may need to make changes so that this code will run in your programming environment.
Send the following commands for this example application:
Command Description
:SENS:FUNC "FRES"
:SENS:FRES:OCOM ON
:SENS:FRES:AZER ON
:SENS:FRES:NPLC 1
•
• Set to measure four-wire resistance
• Enable offset compensation
• Enable auto zero
• Set NPLC to 1
•
Read the resistance value