13 Settings for Stability and Response
Exists the possibility to tune the stability of the display and the degree of reaction of the
balance to the requirements of certain applications or the environmental conditions.
Most measurements can be carried out using default settings, that is, standard mode. In
standard weighing mode, stability and reaction have the same priority. For certain
applications such as e.g. dosage do use the dosing mode (pouring mode). In dosing
mode the reaction degree has the higher priority.
Beside the selection standard / dosing mode the stability of the display and the reaction
degree of the balance can additionally adapted in the menu.
Please note that data processing for greater stability generally slows the response and
processing for higher response reduces stability.
13.1 Stability and reaction settings via „Easy Setting display“
(without invoking menu)
In weighing mode press MENU shortly. The easy setting
display [ ] flashes.
When the display is flashing, set stability or reaction via
UNIT or PRINT as described in the following.
Priority on reaction
Priority on stability
Easy setting display
Operation
Every time the button is hit,
reaction will increase
Every time the button is hit,
stability will increase
The easy setting display flashes for a short time. Inputs are only possible
within this time. Using ON/OFF the flashing Easy setting display can be