15 Check weighing and target mode (check and target weighing)
The check weighing or target mode can be applied for the application
functions (see chpt. 14).
The balance starts in the mode, in which it has been switched off.
15.1 Check weighing
In many cases not the nominal value of the weighed goods is the decisive parameter,
but the deviation from this nominal value. Such applications are for example the weight
check of equivalent packages or the process check of parts in a fabrication process.
The indicators
,
or
in the display show, where the weighed goods are
located within the tolerance range.
These indicators are only functioning in check weighing or in target mode, otherwise
they cannot be seen.
The indicators provide the following information
Condition Grading Indicator
OVR.RNG < sample weight
Beyond tolerance limit
No
indicator
HI.LIM < sample weight OVR.RNG
Upper tolerance limit
LO.LIM sample weight HI.LM
Inside tolerance range
UND.RG sample weight < LO.LIM
Lower tolerance limit
Sample weight < UND.RG
Beyond tolerance limit
No
indicator

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN ABS-N Series and is the answer not in the manual?

KERN ABS-N Series Specifications

General IconGeneral
BrandKERN
ModelABS-N Series
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals