88 ABS-N_ABJ-NM-BA-e-1715
16.4 Data Formats
In the menu item „D.FORM“ four data formats „DF.1-DF.4“ are available.
+ Menu setting, see chpt. 16.3.2, step 7 „Setting data format“.
1. Data format 1 „DF.1“
Example 9.9949 g:
1 20H
Positive weighing value = space character 20H
Negative weighing value = minus sign 2DH
Numeric weighing value is displayed in 8
positions.
Not required positions = space character 20H
Where necessary, Overload is displayed in 2
positions with O L
Weighing unit
12 C/R 0DH
Separator C/R = 0DH, L/F = 0AH
At CR+LF data length will be increased
Printout examples KERN YKB-01N
9.9949g Stable or instable/positive weighing value
-9.9949g Stable or instable/negative weighing value