ABS-N_ABJ-NM-BA-e-1715 89
2. Data format 2 „DF.2“
Example 9.9949 g:
Positive weighing value = space character 20H
Negative weighing value = minus sign 2DH
Stable weighing value = S (stable) 53H
Unstable weighing value = U (unstable) 55H
Numeric weighing value is displayed in 8
positions.
Not required positions = space character 20H
Where necessary, Overload is displayed in 2
positions with O L
Weighing unit
17 C/R 0DH
Separator C/R = 0DH, L/F = 0AH
At CR+LF data length will be increased
Printout examples KERN YKB-01N
S 9.9949 g Stable/positive weighing value
U 9.9949 g Instable/positive weighing value
S -9.9949 g Stable/negative weighing value
U -9.9949 g Instable/negative weighing value