EasyManuals Logo
Home>KERN>Scales>PRS

KERN PRS User Manual

KERN PRS
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
106 PRS/PRJ/ARS/ARJ-BA-def-0622
The calibration of the balance is set in the configuration menu (see Chapter 8.6
“Calibration of the balance”).
Possible types of calibration, depending on the balance model:
External calibration by means of ICM (Intelligent Calibration Mode)
External calibration with freely selectable weight
Internal calibration
Automatic calibration
NOTE
The calibration can be interrupted at any time by pressing the “ON/OFF” key.
6.10.1 External calibration by means of ICM
(only ARS/PRS models)
Depending on the type of balance, calibration weights in steps of 10 g, 50 g, 100 g and
500 g can be used, where the calibration weight must relate to the accuracy of the
balance.
For an external calibration by means of ICM, “SET CALIBRATION MODE EXTERNAL”
must be selected in the configuration menu (see Chapter 8.6 “Calibration functions”).
Switch to “WEIGHING” using the “ ” key
CALIBRATION
Press CAL” until “CALIBRATION” appears.
- - 0000 g
The balance carries out a zero measurement
(0000 g is shown flashing)
- - 2000 g
After the zero measurement has been taken,
the display flashes with the recommended
calibration weight
Place the calibration weight on the pan
- - 2000 g
The display continues to flash
Calibration is complete when the display
stops flashing
+ 2000 g