PRS/PRJ/ARS/ARJ-BA-def-0622 107
6.10.2 External calibration with freely selectable weight
(only ARS/PRS models)
For an external calibration with user-definable weight, “SET CALIBRATION MODE EXT.-
DEF.” must be selected in the configuration menu (see Chapter 8.6 “Calibration
functions”).
Then, the effective value of the calibration weight (DEF. n.nnn g) must be entered with up
to ten times precision when compared with the balance.
NOTE
If calibration is carried out with the free weight, then only this weight may be used.
Then proceed as follows:
• Switch to “WEIGHING” using the “ ” key
CALIBRATION
• Press “CAL” until “CALIBRATION” appears.
- - 0000 g
• The balance carries out a zero measurement
(0000 g is shown flashing)
- - 372 g
• After the zero measurement has been taken,
the display flashes with the previously
entered calibration weight
• Place the calibration weight on the pan
- - 372 g
• The display flashes rapidly
• Calibration is complete when the display
stops flashing (the exact weight is shown)
+372.15 g