158 PRS/PRJ/ARS/ARJ-BA-def-0622
13 Further information
13.1 Notes on Weighing mode
13.1.1 Set Weighing mode: Floating display
The value set for the Floating display defines the period after which each new
measurement is displayed.
When defining this time period, the quality of the balance location is crucial. The stability
control must also be appropriately selected.
Recommended values:
• Optimal balance location: FLOATING DISPLAY 0.04 or
FLOATING DISPLAY 0.08
• Good balance location: FLOATING DISPLAY 0.16
• Critical balance location: FLOATING DISPLAY 0.32
NOTE
The value of the Floating Display is related to the stability control and the balance location.
13.1.2 Set Weighing mode: Stability control
The value set for stability control depends on the quality of the balance location and must
be correctly selected in order to obtain optimal, reproducible results. Select:
• “HIGH STABILITY” at an optimal balance location,
• “MEDIUM STABILITY” at a good balance location or
• “LOW STABILITY” at a critical balance location