166 PRS/PRJ/ARS/ARJ-BA-def-0622
17 Troubleshooting
The balance should be switched off for a short time following any errors in the
programming process and disconnected from the mains supply. It is then necessary to
start the weighing process again from the beginning.
Help:
Fault Possible cause
Weight display is not
illuminated.
• The balance is not switched on.
• The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
• Power supply interrupted.
The weight display changes
continually
• Draught/air movement
• Table/floor vibrations
• The weighing plate is in contact with foreign
matter.
• Electromagnetic fields / static charging (choose
different location/switch off device which is causing
the interference if possible)
The weighing result is
obviously incorrect
• The balance display is not set to zero
• Calibration is no longer correct.
• Great fluctuations in temperature.
• Electromagnetic fields / static charging (choose
different location/switch off device which is causing
the interference if possible)
Switch the balance off if other error messages should appear and then switch on again.
Contact the manufacturer if the error message does not disappear.