3-41
External Device Setup Remark
Draeger Evita 2 dura/ Evita
4/ Evita XL
Protocol: Medibus
Baud Rate: 19200
Parity: even
Stop Bits: 1
Interval: ---( Evita 2 dura)
/
Newport E360 Protocol: Newport
The following information is for
further reference:
Baud Rate: 38400 bps
Word Length: 8 bits
Parity: NONE
Stop Bits: 1
Puritan Bennett 840
Baud Rate: 38400
Word Length: 8 bits
Parity: NONE
/
Hamilton G5protocol
Polling
Protocol: Polling.
The following information is for
further reference:
Baud Rate: 9600 bps
Word Length: 7 bits
Parity: even
Stop Bits: 2
Hamilton C2protocol
Polling
Protocol: Polling.
The following information is for
further reference:
Baud Rate: 9600 bps
Word Length: 7 bits
Parity: even
Stop Bits: 2
Hamilton Galileoprotocol
Polling
Not required.
The following information is for
further reference:
Baud Rate: 9600 bps
Word Length: 7 bits
Parity: even
Stop Bits: 2
Hamilton G5protocol
Block
Protocol: Block.
The following information is for
further reference:
Baud Rate: 38400 bps
Word Length: 8 bits
Parity: none
Stop Bits: 1

Table of Contents

Other manuals for Mindray BeneView T6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T6 and is the answer not in the manual?

Mindray BeneView T6 Specifications

General IconGeneral
BrandMindray
ModelBeneView T6
CategoryMedical Equipment
LanguageEnglish

Related product manuals