Supplementary Reagent Loading (SRL) Module (Optional)
11-5
11.4 Errors Related to the SRL Module
The table below lists the errors might be reported after the SRL module is configured and the
corresponding help messages.
0x00006029
SRL Serial port
Communication error
1. Shut down the analyzer power and SRL module power directly,
and restart them later.
2. If the error still exists, contact our customer service department.
0x00003040
SRL module pressure
out of range
1. SRL module pressure out of range. Sample analysis may be
disabled.
2. Tap the "Remove Error" button to remove this error.
3. If the error still exists, contact our customer service department.
0x00003042
SRL module vacuum out
of range
1. SRL module vacuum out of range. Sample analysis may be
disabled.
2. Tap the "Remove Error" button to remove this error.
3. If the error still exists, contact our customer service department.
0x00004319
Buffer cistern of
supplementary reagent
loading module
abnormal
1. Buffer cistern of supplementary reagent loading module
abnormal. Sample analysis may be disabled.
2. Error cannot be removed. Please contact our customer service
department.